To replace the optical unit, do the following:

1.

Insert the FDT12B Transmitter/ FDR12 Receiver unit into the
hood. Note that the position of the unit is correct, see Figure 34
below.

2.

Connect the ribbon cable connector with the optical unit.

3.

Tighten the three-hexagon socket locking bolts with a 5-mm, T-
handle Allen key.

4.

Replace the round flange firmly. Note the ribbon cable path and
make sure the cable is run along the path safely.

5.

Finally, fasten the outer cover with four screws.

6.

Perform visibility calibration.

Figure 34

Replacing the Optical Units

The following numbers refer to Figure 34 above:

1

=

FDT12B transmitter

2

=

FDR12 receiver

3

=

Photodiode

4

=

Heating elements

5

=

LED