Figure 13

Electronics Enclosure Feedthroughs

The following numbers refer to Figure 13 above:

1

=

Grounding

2

=

DTS14 cable feedthrough

3

=

Temperature sensor (TE)

4

=

Cap (Pg 13.5) of optional opening for the LM11
background luminance meter

5

=

Main power cable

6

=

FDC115 transducer cable feedthrough

7

=

Standard communication cable feedthrough

Communication Cable EMC-shielding

The electronics enclosure has one cable outlet for a cable diameter
from Ø7 to Ø10 mm, which is reserved for a signal or modem cable.
Although the shielding of the cable may be just grounded after cable
inlet, an efficient procedure against RF-interference requires special
care. Ground the cable gland to keep EMI levels within specifications.

For a proper RF-grounding of any jacketed cable, the instructions are
the following:

1.

Lead the signal cable through the cable inlet. If the field cable is
thicker than 10 mm, use a separate signal junction box. See
Figure 14 on page 47.

2.

Strip 80 mm of the cable sheath leaving approximately 40 mm
of the shield.