The format of the AMES command is the following:

AMES Message_number Message_interval

where

Message_number

- The valid range is 0 to 7.
- Selects the corresponding message. Any negative message

number is converted to 0.

- The message number is also the default number for the MES

command and polling.

- If only the message number is given, the previous interval

setting is used.

Message_interval

- Given in multiples of 15 seconds (= the measuring interval).

Therefore intervals 15, 30, 45 ... are valid. Other intervals are
converted to integer multiples of 15 seconds. The maximum
sending interval is 255 seconds (4 minutes 15 seconds).

- Negative or zero interval ignores the automatic sending. This

is used when messages are polled. Refer to section Message
Polling on pa
ge 69 for details.

For example, typing

AMES 1 60

selects message number 1 to be sent once in a minute.

Typing

AMES 0 0

selects message 0, and cancels the automatic sending.

The AMES command without parameters displays the current
selection and it is the following:

AMES

↵↵