Vaisala FD12P User Manual

Page 81

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 81

When no password has been set, the command format is the
following:

CONF

The system response to the CONF command is presented below (The
bold text refers to user actions.)

CONF. PASSWORD (4 CHARS MAX)

UPDATE CONFIGURATION PARAMETERS
UNIT ID (2 CHAR)

( 1)

SET REFERENCE PARAMETERS

(

25.9)

TE

(

127.48) Y

OFFSET
OFFSET REFERENCE UPDATED
ALARM LIMIT 1

(

1000)

ALARM LIMIT 2

(

200) 300

ALARM LIMIT 2

UPDATED

TRANSMITTER CONTAMINATION
LIMITS
WARNING LIMIT

(

1.0) 1.5

WARNING LIMIT UPDATED
ALARM LIMIT

(

5.0)

RECEIVER CONTAMINATION LIMITS
WARNING LIMIT

(

100)

ALARM LIMIT

(

500) 600

ALARM LIMIT UPDATED
ANALOG OUTPUT MODE
0 = LINEAR 1 = LN ( 0)
ANALOG OUTPUT RANGE
MAX VISIBILITY

(

10000)

MIN VISIBILITY

(

50)

END OF CONFIGURATION

The questions asked by the system are described below.

First the system asks for a new password:

CONF. PASSWORD (4 CHARACTERS

MAX)

This question is asked when there is no valid password or the existing
password is updated. If updating is requested by the N parameter and
an empty line is given for an answer, the password is removed.
Otherwise, the user gives a new password to the system.

The system asks the following:

UPDATE CONFIGURATION PARAMETERS

UNIT ID (2 CHAR)( 1)

If the FD12P unit is named by one- or two-character ID codes, the
OPEN and POLLING commands use it as a parameter. The ID code is

Advertising