Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 85

If the LM11 is not connected, the scaling factor should be negative.
Value -1.0 has been set at the factory as the default value. If a positive
value is used, the sensor expects a signal from the LM11.

For an example, see the following:

>blsc
BL SCALE

-1.000

>blsc 10.4

BL SCALE

10.400

If a day/night switch is connected to the serial line control input on the
FDP12 processor board, the FD12P can read the switch state and
report it as a background luminance value of 1 (day) or 0 (night). The
FD12P firmware will read the switch if the background luminance
scaling factor is set to 0.

Maintenance Commands

The maintenance commands are listed in Table 21 below.

Table 21

Maintenance Commands

Command

Description

STA

Displays status.

CAL

Calibrator_frequency Calibration.

TCAL

Temperature measurement calibration.

CLEAN

Sets clean references.

CHEC

Displays average signal.

FREQ

Displays internal signals.

DRY ON

Sets DRD12 dry offset.

WET ON

Sets DRD wet scale.

AN

CHANNEL

Analog channel (0,1,3,8 ... 15 or ANALOG ID).

STA Command

The STA command displays the results from the built-in test system
as a status message. Message 3 gives the same status message as the
STA command.

When you type

STA

↵↵

the system output is the following: