Protection and Status Control Keys

The Protect and Status keys control the protection functions and status registers of the ac source. Refer to
Chapter 4 of the Programming guide for more information on the status registers.

Press this key to access the protection menu list.

Display

Command Function

PROT:CLEAR

Clear the status registers of all activated protection signals.
The fault causing a signal must be corrected or removed
before the register can be cleared.

CURR:PROT <char>

Set overcurrent protection function (ON or OFF).

VOLT:PROT <value>

Set the overvoltage protection level

1

DELAY <value>

Set the time delay for activating a protection fault after
programming the output

Press this key to access the status menu list. Note that in the following list, commands
ending in ? clear the registers when they are read. For this reason the registers are read
only after you press , not when you scroll to the command

Display

Command Function

*CLS

Executes the clear status (*CLS) command

STATUS:PRESET

Execute the STATus:PRESet command

*ESR? <value>

Return Event Status register value

*STB <value>

Return Status Byte register value

OPER:EVEN? <value>

Return STAT:OPER:EVENT? value

OPER:COND <value>

Return STAT:OPER:COND? value

QUES:EVEN? <value>

Return STAT:QUES:EVENT? value

QUES:COND <value>

Return STAT:QUES:COND? value

Notes:

1

Programmed in peak volts. (Other voltage parameters are programmed in rms volts)

value = a numeric value
char = a character string parameter
and scroll through the command list.
and scroll through the parameter list.

Protect

Enter

Status

Shift

q

p
¯

°