Verification and Calibration - B

65

Calibrating and Entering Current Calibration Values

Action

Display

20.

Connect the appropriate current shunt and load resistor as shown in figure B-
1. Connect the DVM (ac rms mode) across the current shunt.

If you are calibrating a 3-phase source, make sure that the default phase 1
(

φ

1) annunciator is lit. Press

Phase Select

to select a different phase.

21.

Press

Shift Calibration,

scroll to CAL CURR AC, and press

Enter

.

CAL:CURR:AC

22.

Press

Shift Calibration

, scroll to CAL LEV P1, and press

Enter

.

CAL:LEV P1

23.

Press

Shift Calibration

and scroll to CAL DATA. Wait for the DVM

reading to stabilize. Read the DVM and calculate the first current value
(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad
to enter the rms current value.

CAL:DATA 0.00

24.

Press

Shift Calibration

, scroll to CAL LEV P1, use

° to scroll to the P2

parameter, and press

Enter

.

CAL:LEV P2

25.

Press

Shift Calibration

and scroll to CAL DATA. Wait for the DVM

reading to stabilize. Read the DVM and calculate the second current value
(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad
to enter the rms current value.

CAL:DATA 0.00

26.

If you are calibrating a 3-phase source, press

Phase Select

to select the

next phase and repeat steps 20 through 25 for phases 2 and 3.
The phase annunciators on the front panel indicate which phase is active.

Calibrating and entering rms current measurement values (Agilent 6843A only)

27.

Press

Shift Calibration,

scroll to CAL CURR MEAS, and press

Enter

.

CAL:CURR:MEAS

28.

Press

Shift Calibration

, scroll to CAL LEV P1, and press

Enter

.

CAL:LEV P1

29.

Press

Shift Calibration

and scroll to CAL DATA. Wait for the DVM

reading to stabilize. Read the DVM and calculate the first current value
(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad
to enter the rms current value.

CAL:DATA 0.00

30.

Press

Shift Calibration

, scroll to CAL LEV P1 command, use

° to scroll

to the P2 parameter, and press

Enter

.

CAL:LEV P2

31.

Press

Shift Calibration

and scroll to the CAL DATA. Wait for the DVM

reading to stabilize. Read the DVM and calculate the second current value
(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad
to enter the rms current value.

The ac source is now holding the new current calibration constants in RAM

CAL:DATA 0.00