background image

4 - Front Panel Operation

54

10 - Switching Between Single- and Three-phase Operation (Agilent 6834B only)

When shipped from the factory, the Agilent 6834B ac source is configured for three-phase operation.
You can configure the Agilent 6834B for single-phase operation. This increases the available output
power for phase 1 from 1.5 kVA to 4.5 kVA.

NOTE:

When you configure the Agilent 6834B for single-phase operation, the phase 2 and phase
3 output terminals are internally connected to the phase 1 output terminal. This
distributes the increased available output power across the three output terminals.

Additionally, switching the Agilent 6834B between single-phase and three-phase operation causes the
following actions:

it disables all outputs

it reconfigures the current readback and programming calibration constants

it returns all lists and *RCL states to their factory default states

it reboots the unit, after which the newly selected number of phases becomes available

NOTE:

This may require you to reprogram the lists and recall states each time the outputs are
switched.

Action

Display

1.

On the System keypad, press

Address

.

ADDRESS 5

2.

Press

ô to scroll through the address menu until you get to the NOUTPUTS

command.

NOUTPUTS 3

3

Use the

¯ and ° keys to select the number of output phases; either 1 (single-phase)

or 3 (three-phase).

NOUTPUTS 3
NOUTPUTS 1