B - Verification and Calibration

64

Calibrating and Entering Voltage Calibration Values

Action

Display

4.

Connect the DVM (ac volts mode) directly to the ac source via the ratio
transformer shown in Figure B-1. Do not connect the load resistor or
current shunt.

5.

Press

Shift Calibration

, scroll to CAL VOLT AC, and press

Enter

.

CAL:VOLT:AC

6.

Press

Shift Calibration

, scroll to CAL LEV P1, and press

Enter

to select

the first calibration point.

CAL:LEV P1

7.

Press

Shift Calibration,

scroll to CAL DATA, and use the Entry keypad

to enter the voltage value displayed on the DVM. (Multiply the DVM
readings by the transformer ratio if a ratio transformer is used.)

CAL:DATA 0.00

8.

Press

Shift Calibration,

scroll to CAL LEV P1, use

° to scroll to the P2

parameter, and press

Enter

. This selects the second calibration point.

CAL:LEV P2

9.

Press

Shift Calibration,

scroll to CAL DATA, and use the Entry keypad

to enter the voltage value displayed on the DVM. (Multiply the DVM
readings by the transformer ratio if a ratio transformer is used.)

CAL:DATA 0.00

10.

Press

Shift Calibration

, scroll to CAL LEV P1, use

° to scroll to the P3

parameter, and press

Enter

.

CAL:LEV P3

11.

Press

Shift Calibration,

scroll to CAL DATA, and use the Entry keypad

to enter the voltage value displayed on the DVM. (Multiply the DVM
readings by the transformer ratio if a ratio transformer is used.)

CAL:DATA 0.00

12.

Press

Shift Calibration

, scroll to CAL LEV P1, use

° to scroll to the P4

parameter, and press

Enter

.

CAL:LEV P4

13.

Press

Shift Calibration,

scroll to CAL DATA, and use the Entry keypad

to enter the voltage value displayed on the DVM. (Multiply the DVM
readings by the transformer ratio if a ratio transformer is used.)

The ac source is now holding the new voltage calibration constants in RAM

CAL:DATA 0.00

Calibrating and entering realtime voltage calibration values (Agilent 6843A
only)

14.

Press

Shift Calibration,

scroll to CAL VOLT RTIM, and press

Enter

.

CAL:VOLT:RTIM

15.

Press

Shift Calibration

, scroll to CAL LEV P1, and press

Enter

.

CAL:LEV P1

16.

Press

Shift Calibration,

scroll to CAL DATA, and use the Entry keypad

to enter the voltage value displayed on the DVM. (Multiply the DVM
readings by the transformer ratio if a ratio transformer is used.)

CAL:DATA 0.00

Calibrating the OVP trip point

17.

Press

Shift Calibration,

scroll to CAL VOLT PROT, and press

Enter

.

CAL:VOLT:PROT

18.

Wait for the ac source to compute the OVP calibration constant. The
display returns to Meter mode when the OVP calculation is complete.

The ac source is now holding the new OVP calibration constants in RAM

19.

If you are calibrating a 3-phase source, press

Phase Select

to select the

next phase and repeat steps 4 through 13 and 17 and 18 for phases 2 and 3.
The phase annunciators on the front panel indicate which phase is active.