Meter display keys, Output control keys – Agilent Technologies 6814B User Manual

Page 49

Advertising
background image

Front Panel Operation - 4

49

On three-phase ac sources, phase 1 is normally offset by 0

°

from the internal phase reference while

phase 2 and phase 3 are offset by 240

°

and 120

°

respectively. Therefore, synchronized transient events

will occur at the phase angle programmed for the phase 1 output, but at different phase angles on the
phase 2 and phase 3 outputs.

This is illustrated in the following figure, where the transient occurs at the 90

°

angle programmed for the

phase 1 output, but at 330

°

and 210

°

for the phase 2 and 3 outputs, because of the default offsets for

these outputs.

4

02

Trigger

VOLT T level

VOLT level

VOLT T level

VOLT level

VOLT T level

VOLT level

03

01

90

0

210

0

330

0

phase = 0

phase = 120

phase = 240

Figure 4-8. Phase Synchronization on Three-phase Sources

Example

Display

m

This example synchronizes a change at 90

°

on phase 1. First, access the

Voltage menu and program the immediate and triggered voltage levels,
followed by the voltage transient mode. Check that the

φ

1,

φ

2, and

φ

3

annunciators are all lit, indicating that commands will be sent to all three
phases. If not, press

Phase Select

until all three annunciators are lit.

Press

Trigger Control

. Press

ô until you access the sync source command.

On the Entry keypad, press

° once to obtain PHASE. Press

Enter

.

Access the Trigger Control menu again and press

ô to access the sync phase

reference parameter. On the Entry keypad, program a 90

°

phase reference by

entering

9 0 Enter

.

Then press

Trigger Control

and

Enter

, followed by

Shift Trigger

.

VOLT 120

VOLT:T 150

VOLT:M STEP

SYNC:SOUR PHASE

SYNC:PHAS 90

INIT:IMMED