Procedure for three-phase ac sources 42 – Agilent Technologies 6814B User Manual

Page 42

Advertising
background image

4 - Front Panel Operation

42

Procedure for Three-Phase AC Sources

If you have a three-phase ac source, you can set the rms voltage of all three phases identically, or set each
one differently. This is controlled via the Phase Select Menu on the Function keypad.

The following example shows how you can set the phase 1 output to 120 Vrms, phase 2 to 180 Vrms,
and phase 3 to 235 Vrms.

Action

Display

1.

On the Function keypad, press

Voltage

. Note that in the power-on default state, the

φ

1,

φ

2, and

φ

3 annunciators are all lit, indicating that any command sent to the

instrument will be sent to all three phases.

VOLT 0

2.

Scroll through the phase selections by pressing

Phase Select

Pressing

Phase Select

moves you through phases 1, 2, and 3, as indicated by the

phase annunciators. Pressing it once more returns you to the first state with all three
phase annunciators on.

VOLT 0

3.

While in the Voltage menu press

Phase Select

to access Phase 1. On the Entry pad

press

1 2 0 Enter

.

VOLT 120

4.

Access the Voltage menu again and press

Phase Select

once to access Phase 2. On

the Entry pad press

1 8 0 Enter

.

VOLT 180

5.

Access the Voltage menu again and press

Phase Select

once to access Phase 3. On

the Entry pad press

2 3 5 Enter

.

VOLT 235

6.

Press

Output On/Off

to enable the output. The Dis annunciator will go off,

indicating that the voltages are now applied to the output terminals.

235 V 60 Hz

To verify the output, you can measure it as follows:

7.

Note that the

φ

3 annunciator is on, indicating that you are monitoring the last phase

you selected, which was Phase 3.

235 V 60 Hz

8.

Press

Phase Select

once. The

φ

1 annunciator will light, indicating that you are

now monitoring Phase 1.

120 V 60 Hz

9.

Press

Phase Select

once. The

φ

2 annunciator will light, indicating that you are

now monitoring Phase 2.

180 V 60 Hz

10.

Press

Phase Select

once. The

φ

3 annunciator will light, indicating that you are

now monitoring Phase 3 again.

235 V 60 Hz

NOTE:

The Meter menu does not go to the all-phases state present in the Phase Select menu
because the front panel can display only one phase at a time. Refer to the ac source
Programming Guide on how to return simultaneous measurements from all three phases.