Scrolling keys – Agilent Technologies 6814B User Manual

Page 46

Advertising
background image

4 - Front Panel Operation

46

List Transient

Lists are the most flexible means of generating multiple or synchronized transient outputs. The following
figure shows a voltage output generated from a list. The output shown represents three different ac
voltage pulses (160 volts for 33 milliseconds, 120 volts for 83 milliseconds, and 80 volts for 150
milliseconds) separated by 67-millisecond, 0-volt intervals.

The list specifies the pulses as three voltage points (point 0, 2, and 4), each with its corresponding dwell
point. The intervals are three zero-voltage points (point 1, 3, and 5) of equal intervals. The count
parameter causes the list to execute twice when started by a single trigger.

NOTE:

From the Output Menu, execute the *RST command to reset the ac source. This is
necessary because any previously programmed functions remain in effect until cleared.

Figure 4-6. List Transients

Action

Display

1.

Press

Voltage

to access the Voltage Menu. Then press

ô to access the voltage mode

command.

VOLT:M

FIXED

2.

On the Entry keypad, press

¯ or ° to scroll through the mode parameters to obtain LIST

and press

Enter

.

VOLT:M LIST

3.

Access the List Menu by pressing

Shift List

. The first menu command is the count.

From the Entry keypad, change the list count from the default (1) to 2. Press

Enter

.

COUNT 2

4.

Access the List menu again and press

ô until you access the dwell time. This specifies

the "on" time for each voltage point, which is effectively the output pulse width. The first
dwell point (0) appears in the display. On the Entry keypad, press

. 0 3 3

and

Enter

.

DWEL 0 .033

5.

Pressing the Enter key automatically advances to the step in the list. Enter the following
values for dwell list points 1 through 5: .067, .083, .067, .150, .067. Press

Enter

to

enter each value. When you finish, you will be at point 6, which is the end of the list.

Note: Press

Shift



Index

or

Shift

ô

Index

to access and edit any list point.

DWEL 1 .067
DWEL 2 .083
DWEL 3 .067
DWEL 4 .150
DWEL 5 .067

DWEL 6 EOL

160 Vrms

Trigger

0

1

2

3

4

5

List Count = 1

List Count = 2