Front Panel Operation - 4

51

Example

Display

j

This example uses the immediate slew rate. First, access the Voltage menu
and press

ô until you access the mode command. On the Entry keypad, press

° to obtain FIXED. Press

Enter

.

Access the voltage menu and press

ô until you access the slew command. On

the Entry keypad, press

5 0 Enter

to program a slew rate of 50 volts/second.

Whenever a new immediate voltage value is entered, the output will slew to
the new level at 50 volts/second.

VOLT:M FIXED

SLEW 50

ô

Step mode uses the triggered slew rate. First, access the Voltage menu,
program the immediate and triggered voltage levels, and set the slew mode to
STEP.

Access the Voltage menu and press

ô to access the immediate slew

command. On the Entry keypad, enter a value that equals infinity.

Access the Voltage menu and press

ô until you access the triggered slew

command. On the Entry keypad, enter a value such as

5 0 Enter

, which sets

the triggered slew rate to 50 volts/second.

Then press

Trigger Control

and

Enter

, followed by

Shift Trigger

.

After the trigger has been sent, in step mode, the triggered value becomes the
new immediate value.

VOLT 120

VOLT:T 150

SLEW:M STEP

SLEW: 9.9+E37

SLEW:T 50

INIT:IMMED

í

Pulse mode uses the triggered slew rate at the leading edge of the pulse, and
the immediate slew rate at the trailing edge of the pulse. First, access the
Voltage menu, program the immediate and triggered voltage levels, and set
the slew mode to PULSE.

Access the Pulse menu and program the pulse count, duty cycle, and pulse
period.

Access the Voltage menu and press

ô to access the immediate slew

command. On the Entry keypad, enter a value that equals infinity.

Access the Voltage menu and press

ô until you access the triggered slew

command. On the Entry keypad, enter a value such as

5 0 Enter

, which sets

the triggered slew rate to 50 volts/second.

Then press

Trigger Control

and

Enter

, followed by

Shift Trigger

.

VOLT 120

VOLT:T 150

SLEW:M PULSE

COUNT 2

DCYCLE 33

PER .0166

SLEW: 9.9+E37

SLEW:T 50

INIT:IMMED

÷

When the voltage slew mode is set to LIST, the slew rates are set by the
values in the voltage slew list. Refer to the List Transient example for more
information on how to program lists. You must program the voltage values
and dwell times as explained in that example. You must also program a slew
rate for each point in the list (even if it is 9.9+E37).

NOTE:

When specifying a dwell time, you must take the slew time into consideration. If the
dwell time at any given list point is less than the slew time at the same point, the voltage
will never reach its programmed level before the next list point becomes active.