Turn-on checkout – Agilent Technologies 6814B User Manual

Page 27

Advertising
background image

Turn-On Checkout - 3

27

Make sure that the unit is turned off, and make the following connections to the output. If you are
verifying a single phase source, you only need to connect one bulb.

Figure 3-1.Verification Connections

Procedure

Display

Explanation

1.

Turn the unit on.

Meter mode

Meter mode is active and the Dis annunciator should be
on.

2.

Press the

Voltage

key.

VOLT 0.00

Display indicates the default settings.
If you are verifying a three phase source, all phase
annunciators should be on.

3.

Press

Phase Select

until only

φ

1 is lit.

VOLT 0.00

Single phase units use only phase 1.

4.

Press

1, 2, 0, Enter

VOLT 120

Programs the output on phase 1 to 120 Vrms. After the
value is entered, the display returns to Meter mode,
which indicates that no voltage is applied to the output.

5.

Press

Output On/Off

120 V 60 Hz

Turns the output on and applies 120 volts to the phase 1
light bulb. The Dis annunciator should be off and CV
should be on.

If you are verifying three phase sources, repeat steps 2 through 4 for phase 2 and phase 3. All three
light bulbs should be lit when this action is complete.

Continue with step 6 of you are verifying three phase sources. Otherwise skip to step #10.

6.

Press the

Voltage

key.

VOLT 120

7.

Press

Phase Select

until all phase
annunciators are lit

VOLT 120

8.

Press

0, Enter

VOLT 0

Turns off all light bulbs connected to the three-phase
source.

9.

Press

Voltage 1, 2, 0,

Enter

VOLT 120

Turns on all light bulbs connected to the three-phase
source.

10.

Press

Protect

PROT:CLEAR

Display accesses the protection menu list.