Verification and calibration, B. verification and calibration 59, Introduction 59 – Agilent Technologies 6814B User Manual

Page 59: Equipment required 59, Introduction, Equipment required

Advertising
background image

59

B

Verification and Calibration

Introduction

This appendix includes verification and calibration procedures for the Agilent 6814B/6834B/6843A AC
Power Solutions. Instructions are given for performing the procedures either from the front panel or from
a controller over the GPIB.

The verification procedures do not check all the operating parameters, but verify that the ac source is
performing properly. Performance Tests, which check all the specifications of the ac source, are given in
the applicable ac source Service Manual.

Because the output of the ac source must be enabled during verification or calibration, proceed with
caution, since voltages and currents will be active at the output terminals.

Important

Perform the verification procedures before calibrating your ac source. If the ac source
passes the verification procedures, the unit is operating within its calibration limits and
does not need to be recalibrated.

WARNING

LETHAL VOLTAGES. Ac sources can supply 424 V peak at their output. DEATH
on contact may result if the output terminals or circuits connected to the output are
touched when power is applied. These procedures must be performed by a qualified
electronics technician or engineer trained on this equipment.

Equipment Required

The equipment listed in the following table, or the equivalent to this equipment, is required for
verification and calibration.

Table B-1. Equipment Required

Equipment

Characteristics

Recommended Model

Digital Voltmeter

Resolution: 10 nV @ 1 V
Readout: 8.5 digits
Accuracy: >20 ppm

Agilent 3458A

Current Monitor

1

0.01

,

±

200 ppm, (Agilent 6814B/6834B)

0.001

,

±

200 ppm, (Agilent 6843A)

Guildline 7320/0.01
Norma Gerts Instruments

Ratio Transformer

2

30:1 ratio, 50 ppm, 45 Hz to 1 kHz

Load Resistors

2

15

, > 1.5 kW (Agilent 6814B/6834B)

1

5

, > 4.8 kW non-inductive (Agilent 6843A)

Impedance Resistor

1

, 100 Watts min. (Agilent 6843A only)

GPIB Controller

Full GPIB capabilities

HP Series 200/300 or equivalent

1

The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the

load leads and connections. Connect the voltmeter directly to these current-monitoring terminals.

2

A ratio transformer is required only when verifying output voltage readback to MIL-STD-45662A 4:1 test

equipment ratio requirements.