Front panel calibration, Front panel calibration menu 63, Front panel calibration 63 – Agilent Technologies 6814B User Manual

Page 63: Enable calibration mode 63, Front panel calibration menu, Enable calibration mode

Advertising
background image

Verification and Calibration - B

63

Front Panel Calibration Menu

The Entry keypad is used for calibration functions.

Press this key to access the calibration menu.

Display

Command Function

CAL ON <value>

Turns calibration mode on when the correct password
value is entered.

CAL OFF

Turns calibration mode off

CAL:LEV <char>

Advance to next step in sequence (P1, P2, P3, or P4).

CAL:DATA <value>

Input a calibration measurement.

CAL:VOLT:AC

Begin ac voltage calibration sequence

CAL:VOLT:PROT

Begin voltage protection calibration

CAL:CURR:AC

Begin ac current calibration sequence

CAL:CURR:MEAS

Begin current measurement calibration sequence

CAL:IMP

Begin output impedance calibration sequence

CAL:VOLT:RTIM

1

Begin realtime voltage calibration

CAL:SAVE

Saves the calibration constants in non-volatile memory.

CAL:PASS <value>

Set new calibration password.

Notes:

1

Agilent 6843A only

value = a numeric value
char = a character string parameter
Use and to scroll through the command list.
Use and to scroll through the parameter list.

Front Panel Calibration

WARNING

LETHAL VOLTAGES. Ac sources can supply 425 V peak at their output. DEATH
on contact may result if the output terminals or circuits connected to the output are
touched when power is applied. These procedures must be performed by a qualified
electronics technician or engineer trained on this equipment.

The following procedures assume you understand how to operate front panel keys (see Chapter 4).

Enable Calibration Mode

Action

Display

1.

Reset the unit by selecting

Shift

,

Output

, and pressing

Enter

.

If you are calibration a 3-phase source, make sure that the default phase 1
(

φ

1) annunciator is lit. Press

Phase Select

to select a different phase.

*RST

2.

Press

Shift Calibration

, scroll to CAL ON, and press

Enter

.

CAL ON 0.0

3.

Enter the calibration password from Entry keypad and press

Enter

. If the

password is correct the Cal annunciator will come on.

If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed. (see the Service Manual.)

If the password is incorrect, an error occurs. If the active password is lost,
the calibration function can be recovered by setting an internal switch that
defeats password protection. (see the Service Manual.)

CAL DENIED

OUT OF RANGE

Shift

Cal

q

p
Л

М