B - Verification and Calibration

66

Calibrating the Output Impedance (Agilent 6843A only)

NOTE:

The output voltage and output current must be calibrated before the output impedance
can be calibrated.

Action

Display

32.

Connect only the output impedance resistor across the output of the ac source.
Do not connect any other equipment.

33.

Press

Shift Calibration,

scroll to the CAL IMP command, and press

Enter

.

CAL:IMP

34.

Wait for the ac source to compute the output impedance calibration constant.
The display returns to Meter mode when the calculation is complete.

The ac source is now holding the new impedance calibration constants in RAM

Saving the Calibration Constants

CAUTION:

Storing calibration constants overwrites the existing ones in non-volatile memory. If you
are not sure you want to permanently store the new constants, omit this step. The ac
source calibration will then remain unchanged.

Action

Display

35.

Press

Shift Calibration,

scroll to CAL SAVE, and press

Enter

.

CAL:SAVE

36.

Press

Shift Calibration,

select CAL OFF, and press

Enter

to exit Calibration

mode. *RST and *RCL will also set the calibration state to OFF.

CAL OFF

Changing the Calibration Password

The factory default password is 0. You can change the password when the ac source is in calibration
mode (which requires you to enter the existing password). Proceed as follows:

Action

Display

1.

Begin by pressing

Shift Calibration

and scrolling to the CAL ON command.

CAL ON 0.0

2.

Enter the existing password from Entry keypad and press

Enter

3.

Press

Shift Calibration

and scroll to the CAL PASS command.

CAL:PASS 0

4.

Enter the new password from the keypad. You can use any number with up to
six digits and an optional decimal point. If you want the calibration function to
operate without requiring any password, change the password to 0 (zero).

NOTE:

If you want the calibration function to operate without requiring any password, change
the password to 0 (zero).