Verification and Calibration - B

67

Calibration Error Messages

Errors that can occur during calibration are shown in the following table.

Table B-3. GPIB Calibration Error Messages

Error

Meaning

401

CAL switch prevents calibration (This is a hardware disable, see the ac source Service Manual.)

402

CAL password is incorrect

403

CAL not enabled

404

Computed readback cal constants are incorrect

405

Computed programming cal constants are incorrect

406

Incorrect sequence of calibration commands

Calibration Over the GPIB

You can calibrate the ac source by using SCPI commands within your controller programming
statements. Be sure you are familiar with calibration from the front panel before you calibrate from a
controller. Each front panel calibration command has a corresponding SCPI command.

The SCPI calibration commands are explained in Chapter 3 of the ac source Programming Guide.
Calibration error messages that can occur during GPIB calibration are shown in table B-3.

Agilent Calibration Program Listing

Figure B-2 lists the calibration program. This program can be run on any controller operating under
Agilent BASIC. The assumed power supply address is 705 and calibration password is 0. If required,
change these parameters in the appropriate statements.