9. Maintenance and Troubleshooting

These guidelines apply to several different Campbell Scientific weather

stations.

9.1 Maintenance

Proper maintenance of weather station components is essential to obtain

accurate data. Equipment must be in good operating condition, which requires

a program of regular inspection and maintenance. Routine and simple

maintenance can be accomplished by the person in charge of the weather

station. More difficult maintenance such as sensor calibration, sensor

performance testing (i.e., bearing torque), and sensor component replacement,

generally requires a skilled technician, or that the instrument be sent to

Campbell Scientific or the manufacturer.

A station log should be maintained for each weather station that includes serial

numbers, dates that the site was visited, and maintenance that was performed.

9.1.1 Instrumentation Maintenance

The instrumentation requires a minimum of routine maintenance. A few

preventative maintenance steps will optimize battery life and decrease the

chances of datalogger failure.

9.1.2 Batteries

The CRBasic Battery instruction can be used to measure the CR1000’s battery

voltage. By recording battery voltage the user can determine how long a fresh

set of batteries will last (see the Installation Section of the datalogger

Operator’s Manual for cold temperature effects on alkaline batteries). Short

Cut automatically program the weather station to measure battery voltage.

When alkaline batteries are used, the battery voltage should not be allowed to

drop below 9.6 Vdc before replacement. Where CR10 or 21X dataloggers are

used in the instrumentation, an external battery must be used to maintain power

to the datalogger when changing batteries, otherwise the clock, program, and

data will be lost (refer to the Installation Section of the datalogger’s Operator’s

Manual for details). When not in use, remove the eight cells to eliminate

potential corrosion of the contact points, and store in a cool dry place.

Rechargeable power supplies should be connected to an AC transformer or

unregulated solar panel at all times. The charge indicating diode should be

“ON” when voltage to the charging circuitry is present. Be aware of battery

voltage that consistently decreases over time, which indicates a failure in the

charging circuitry.

9.1.3 Desiccant

In standard weather stations, a humidity indicator card is provided with the

enclosure. A small RH sensor (pn 10162) can be purchased separately to

record the RH inside the enclosure. Change the desiccant when either the card

or the sensor read about 35% RH.

Desiccant may be ordered through Campbell Scientific (DSC 20/4).