4 sensor maintenance, Sensor maintenance – Campbell Scientific UT20 and UT30 Tower-based Weather Stations User Manual

Page 67

Advertising
background image

UT20 and UT30 Tower-based Weather Stations

Desiccant packs inside of the dataloggers do not require replacement under

normal conditions.

9.1.4 Sensor Maintenance

Sensor maintenance should be performed at regular intervals, depending on the

desired accuracy and the conditions of use. A suggested maintenance schedule is

outlined below.

1 week

Check the pyranometer for level and contamination. Gently clean, if

needed.

Visually inspect the wind sensors and radiation shield.

1 month

Check the rain gage funnel for debris and level.

Do a visual/audio inspection of the anemometer at low wind speeds.

Check the filter of the temperature/humidity sensor for contamination.

General Maintenance

An occasional cleaning of the glass on the solar panel will improve its

efficiency.

Check sensor leads and cables for cracking, deterioration, proper routing,

and strain relief.

Check the tripod or tower for structural damage, proper alignment, and for

level/plumb.

6 months

Clean the temperature/humidity sensor.

Clean the Gill Radiation Shield.

1 year

Replace anemometer bearings.

Calibrate the rain gage.

Calibrate the HMP45C probe.

Check calibration of HMP50 RH Probe; replace RH chip if necessary.