Select the test options for voltage fluctuations – Agilent Technologies EN 61000-3-2 User Manual

Page 23

Advertising
background image

Setting Up the Test - 3

23

Test Termination Options

Termination options let you select the conditions that will terminate the test. They are selected in the
Termination window.

Note

The following conditions terminate any test automatically if the Agilent 6800-Series AC
Power Source/Analyzer operating limits are exceeded: Rail fault condition,
Overtemperature, Overvoltage, and SOA. Refer to the applicable ac source User’s
Guide for an explanation of these conditions.

ñ Select termination on errors. One error will terminate the test when this is selected:

UNR - when the output becomes unregulated.

ñ Select termination when the current limit is exceeded. The two current limit conditions that will

terminate the test when this occurs are: CL - when the output goes into current limit mode, and
CP - when the output current limit protection has tripped.

ñ Select termination upon exceeding a specified number of failures. When testing fluctuating

harmonics and this box is checked, you can specify failures:

of any type,
with a 2.5 minute window
without a 2.5 minute window.

ñ Select termination on exceeding a specified number of watch events. Each event that satisfies the

selected criteria is counted as one event occurrence.

Advertising