D class determination, Class a device selected, System:configure – Agilent Technologies EN 61000-3-2 User Manual

Page 93

Advertising
background image

Command Summary - C

93

SYSTem:CONFigure

This command sets the overall operating mode for the Agilent 6800-Series AC Power Source/Analyzers.
The choices are normal mode, which causes the product to operate as a standard ac source, or IEC mode,
which modifies the basic behavior of the transient and measurement systems to facilitate IEC
measurements. SYSTem:CONFigure has a variety of global consequences that are summarized below:

NORMAL MODE

IEC MODE

Base Sampling Rate

39.920792 kHz

38.400000 kHz

Output Frequency

DC - 1000 Hz

50 Hz & 60 Hz Only

Freq/Window/Fs Mode

Independent

Coupled

Transient System

FIXEd/STEP/PULSe/LIST
Modes

FIXEd Mode Only

Slew Operation

AC & DC Voltage; Frequency

AC Voltage Only

MEASure:ARRay:CURRent
:HARM?

DC, Fundamental, and
Harmonics to 50th

Fundamental and Harmonics
to 40th plus RMS Current,
RMS Voltage & Real Power

CALCulate:SMOothing

Not/Available

1.5 Second Smoothing on/off

MEASure:ARRay:VOLTage
:FLUCtuations

Not/Available

FLICker | PST | ALL?

CALCulate:INTegral:TIME

Not/Available

1 | 5 | 10 | 15 MINutes

Transmission of a SYSTem:CONFigure command implies ABORt and terminates any transient or
measurement actions previously initiated.

Command Syntax

SYSTem:CONFigure <CRD>

Parameters

NORMal | IEC

*RST Value

NORMal

Examples

SYST:CONF NORM

Query Syntax

SYSTem:CONFigure?

Returned Parameters

<CRD>

Related Commands

ABORt MEAS:ARR:CURR MEAS:ARR:VOLT CALC:SMO
CALC:INT SENS:WIND