3 - Setting Up the Test

26

Test Termination Options

Note

The following conditions terminate any test automatically if the Agilent 6800-Series AC
Power Source/Analyzer operating limits are exceeded: Rail fault condition, Overtemperature,
Overvoltage, and SOA. Refer to the ac source User’s Guide for more information.

ñ Select termination on errors. One error will terminate the test when this is selected:

UNR - when the output becomes unregulated.

ñ Select termination when the current limit is exceeded. Two current limit conditions will

terminate the test when this occurs: CL - when the output goes into current limit mode; CP -
when the output current limit protection has tripped.

ñ Select termination on exceeding a specified number of Pst failures
ñ Select termination on exceeding a specified number of Rms voltage failures. You can specify

Rms voltage failures: of any type

Dmax failures
Dc failures
Dt failures.

ñ Select termination on exceeding a specified number of watch events. Each event that satisfies the

selected criteria is counted as one event occurrence.

If termination On Watch Events is enabled, you can select the following Watch Items criteria:

ñ Pst values greater than the specified perceptibility units
ñ Instantaneous Flicker values greater than the specified perceptibility units