Command Summary - C

87

MEASure:ARRay:CURRent:HARMonic?

This query returns an array of current harmonic magnitudes. Operation of the query is modified by the
SYSTem:CONF command (see summary table under SYSTem:CONFigure). The parameter specifies the
number of harmonic arrays to be returned in response to the query. If SYSTem:CONFigure specifies
NORMal operation, the parameter is ignored (ie it is forced to 1). If SYSTem:CONFigure specifies IEC
operation, the SOURce:FREQuency and SENSe:WINDow commands are coupled to modify operation of
the measurement underlying the query as shown in the following table:

SOURce:FREQ

SENSe:WINDow

Sample Rate

Acquisition
Window

Acquisition
Overlap

50 Hz

RECTangular

12.8 KHz

320 ms

None

50 Hz

HANNing

8.533 KHz

480 ms

50%

60 Hz

RECTangular

15.360 KHz

266.7 ms

None

60 Hz

HANNing

7.680 KHz

533.3 ms

50%

SYSTem:CONFigure also impacts availability of the RMS Current, RMS Voltage, Real Power values. If
SYSTem:CONFigure is set to NORMal, these values are not available. If SYSTem:CONFigure is set to
IEC, the values are returned with the harmonic data. The integration time for these values equals the
acquisition window period.

IEC mode operation conforms to IEC and EN requirements for compliance testing of harmonic currents
(EN 60555 Part 2 and related regulations). The ac source will accept parameters in the range shown
below, however, values greater or equal to (2^31)-1 will be interpreted as infinity. Record numbering
begins with one. The figure below defines the structure of the data returned by this query:

Fundamental

##

Ж

Ж

Ж

"###

40 harmonic values

###

Ж

Ж

40th Harmonic

##D

Ж

Ж

Rms Current

##########################З

Ж

Rms Voltage

##########################Ç

"##

repeat <n> times

Real Power

##########################З

Ж

Record Number

##########################З

Ж

Error Code

##########################D

Command Syntax

MEASure:ARRay:CURRent:HARMonic? <NRf+>

Parameters

1 to 9.9E37

Examples

MEAS:ARR:CURR:HARM? 1024

Related Commands

ABORt SYST:CONF INST:NSEL SENS:WIND

SOUR:FREQ