Short Form Report

The Short Form report includes summary information about the current test. This information fits on one page.
The following is a sample short form report of a Fluctuating Current Harmonics test:

IEC 1000-3-2/EN 61000-3-2 Fluctuating Current Harmonics Test
Date Performed: 06/15/97

Test Executed By: A.C. Testright
Company Name: True Test Laboratories
Device Under Test ID: Model 12345 Modular Power Supply
Test ID: 00123

Approved by: QA
Signature:________________________________________________________ Date:__________________

Final Test Result: FAIL

Settings and Test Conditions Compliant to the Standard: Yes

Test Equipment Used:

Agilent 6842A Harmonic/Flicker Test System with serial number 3440A-00101

Agilent 14761A HFTS software Version: A.00.05

Date Last Calibrated: June 6, 1997

Test Equipment Settings:
------------------------
Line Voltage: 230.00 V

Current Measurement Range: High

Line Frequency: 50 Hz

Measurement Delay: 10.0 seconds

Device Class: D

Fluctuating Harmonics Test Duration: 2.00 minutes

Class Determination Pre-test Duration: 10.00 seconds

Overrides:
----------
Test Limit Source (Power Measurements/Statistics): Maximum
Power Overrides: None
Test Limit Overrides: None

Pre-test Results for Class Determination:
-----------------------------------------

Percent in Envelope: 100.0%

Fundamental Current: 1.343 A

Class D Equipment?: Yes

Real Power: 148.8 W

Voltage THD Out-of-Specification?: No

Power Factor: 0.387

Maximum Power: 148.8 W

Mean Power: 146.7 W

Total Number of Failures:

Total Number of Errors:

-------------------------

-------------------------

4505 Failures

None