C - Command Summary

92

SENSe:CURRent:PREFerence

This command sets the phase reference for current harmonic phase measurements. If VOLTage is
selected, the reference is the fundamental component of the measured output voltage. If CURRent is
selected, the reference is the fundamental component of the measured output current.

Command Syntax

SENSe:CURRent:PREFerence <CRD>

Parameters

VOLTage | CURRent

*RST Value

CURRent

Examples

SENS:CURR:PREF CURR

Query Syntax

SENSe:CURRent:PREF?

Returned Parameters

<CRD>

Related Commands

ABORt MEAS:ARR:CURR:PHAS

SENSe:WINDow

This command sets the window function which is used in harmonic measurements. The choice of
parameters is affected by the SYSTem:CONF command. If NORMal is selected, HANNing, KBESsel,
or RECTangular may be selected. IF IEC mode is selected, only HANNing and RECTangular may be
selected. KBESsel is the preferred window and should be used for most measurements in NORMal
mode. HANNing and RECTangular are available for making harmonic current measurements that
comply with the regulatory requirements.

Command Syntax

SENSe:WINDow [:TYPE] <window>

Parameters

HANNing | KBESsel | RECTangular

*RST Value

KBESsel
(RECTangular is the default setting when the HFTS software is run)

Examples

SENS:WIND RECT

Query Syntax

SENSe:WINDow?

Returned Parameters

<CRD>

Related Commands

MEASure:ARRay:CURRent:HARMonic? SYSTem:CONF
MEASure:CURRent:HARMonic?
MEASure:ARRay:VOLTage:HARMonic?
MEASure:VOLTage:HARMonic?