Instrument operation, Lcd codes – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual

Page 10

Advertising
background image

4. INSTRUMENT

OPERATION


LCD Codes

LCD CODE

EXPLANATION

000

Ready to sample

.000

No hydrogen sulfide reading

.8.8.8 Sensor

saturated-regeneration needed (refer to page 12)

.H.H.H

Sensor regeneration in progress (.H.H.H flashes)

.L.L.L

Re-zero needed (refer to page 13)

.P.P.P

Power cord required or low line power, <100 VAC (or 200 VAC)
(see pages 16 and 17, Changing the Fuse, if .P.P.P remains on after
the cord is connected.)

.LO BAT

Recharge batteries (refer to page 17)

.E.E.E

Same as LO BAT, automatically shuts off

.HL

Very high concentration has been detected. Refer to your safety
policy for additional direction to confirm the concentrations."

DURING SAMPLING

.-

0-25% sensor saturation

.--

25-50% sensor saturation

.---

50-75% sensor saturation

-.---

75-100% sensor saturation

DURING SAMPLING, USING SURVEY MODE

-

Survey sampling (minus sign flashes continuously)

WHEN ZERO IS

DEPRESSED

Adjust to 0 only after sensor regeneration. It is normal for the
display to read H after sampling has started.

0

Zero, ready to sample

H

High, turn Zero potentiometer counterclockwise

L

Low, turn Zero potentiometer clockwise

Page 10 of 50