5. MAINTENANCE

Preventive Maintenance Calendar

To keep the Jerome 631-X operating at peak performance, follow the maintenance schedule below
as a guide. Since maintenance is more a function of application and amount of use rather than
time, your requirements may be different from the listed schedule. Call AZI Customer Service at
800-528-7411, 602-470-1414, or e-mail [email protected] for additional guidance for your
environment and operation.

PART/COMPONENT

MAINTENANCE CYCLE

REFER TO

PAGE

Charge batteries

At least once per month, after 1 month's storage,
or when LO BAT appears

Page 17

Change .25 inch fritware

Weekly or as needed

Page 19

Change internal filters and
tubing

After 6 months of use or as needed

Page 20

Replace zero air filter

1

Annually

Factory calibration

Annually

Page 23

Calibration check

Monthly or as needed

Appendix
A, Page 36

Replace batteries

Annually or as needed.
The battery pack contains NiCad batteries.
Dispose of properly.

Page 21


NOTE:
Install the zero air filter into the instrument's intake during storage.

1

Zero air filters, LFS and LFD scrubber filters contain Resisorb™. Call AZI Customer Service at 800-528-7411,

601-470-1414, or e-mail [email protected] for a copy of the Material Safety Data Sheet. Dispose of all filters
properly.