13. APPENDIX C - INTERNAL DIP SWITCH SETTINGS


The DIP switch identified as SW2 on the
631-X control board is located at the top
center of the board.







CAUTION:

Switch the instrument’s power to OFF before opening the instrument

and changing DIP switch settings.

DIP Switch Settings:

Switch # Position

Function

Comment

1

ON

AUTORANGE

DISABLED (SEE DIP SWITCHES 5 & 6)

1 OFF

AUTORANGE

ENABLED

2 X

NOT

USED

3 X

NOT

USED

4 ON

ZERO DISPLAY:

00-99

4 OFF

ZERO DISPLAY:

L-O-H

Switch #

Function

Comment

5 6

RANGE SAMPLE

TIME

ON

ON

0 (.000 - .099PPM) 30 SECONDS

ON

OFF

1 (0.10 - 0.99 PPM) 25 SECONDS

OFF

ON

2 (01.0 - 09.9 PPM) 16 SECONDS

OFF

OFF

3 (10 - 50 PPM)

13 SECONDS