Operating on AC Power or Generator

• For stationary use, the 631-X may be operated on AC power.

Operating the instrument on AC power at all times eliminates the need for the battery
pack and its necessary maintenance.

The battery may be unplugged or removed completely whenever the instrument is
operating on AC power.

• When a generator is used to power the Jerome 631-X, a high quality line conditioner or

voltage regulator is required to ensure a pure sine wave and regulated voltage is applied to
the instrument. The gold film sensor may be damaged by voltage that varies in amplitude
or by surges, spikes, and/or noise on the power line.

Operating on Internal Battery Power

• For portable use, the 631-X may be operated on Battery power.

When you operate the instrument on battery power, please be aware of the following:

A fully charged battery pack, AZI P/N Z4000-0907, provides power for a minimum of
six (6) hours of operation.

For operating more than six (6) hours, an extra fully charged battery pack is needed.

Complete battery recharging takes 14 hours. Refer to Charging Batteries on page 17.

The 631-X uses a rechargeable Nickel Cadmium (NiCad ) battery. Dispose of worn-
out batteries properly when you are replacing the battery pack.

External battery power

A special version of the Jerome 631-X and a DC Power Kit are available to operate the instrument
from a secondary DC source. The source may be a car/truck battery or a storage cell used in
conjunction with solar panels.

Call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail [email protected] for
additional information.