Replacing the battery pack, Setting the input voltage, 21 if necess – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual

Page 21

Advertising
background image

Replacing the Battery Pack

• Press the power OFF button.

• Unplug the power cord.

• Remove the two (2) side screws from

the intake end of the instrument and
open the case lid.

• Disconnect the battery connector

from the board.

• Loosen the two (2) captive screws

holding the battery bracket and
remove the bracket.

• Remove the old battery pack and

replace with a new battery pack.

• Replace the battery bracket and

tighten the captive screws.

• Connect the new battery connector to

the board.

• Close the case and replace the two (2) side screws.

• Dispose of the old NiCad battery in accordance with state and federal regulations.


Setting the Input Voltage

Instruments are factory set and calibrated to use the power setting requested on the order.
However, the voltage setting is easily changed to use either 110 or 220 VAC.

• Ensure the instrument is turned OFF and

unplugged.

• Locate the voltage selector on the rear of the

instrument.

• Insert a small screwdriver in the voltage

selector slot and turn the selector until the
arrow points toward your setting choice and
a click is heard.

AZI Customer Service 800-528-7411 or 602-470-1414