Optional communications capability, Instrument i/o interface – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual

Page 28

Advertising
background image


Optional Communications Capability

Alarm output

30V DC, 100mA

Data output

1. RS-232 Serial, Baud Rate 1200 for use with data

logger, and/or Jerome communication program.

2. RS-232 Serial data format with 0 & 20mA current

logic levels; Baud Rate 1200 (special industrial
applications) and Analog 20 mA output.

"OPTION BOARD" - See APPENDIX D - JEROME 631-X OPTION BOARD on page 40.

Analog output

0 to 2V or 4 to 20 mA

Auto sample interval

5, 15, 30, or 60 minutes

Auto regeneration interval

6, 24 or 72 hours



Instrument I/O Interface

The 631-X I/O port (25 pin D-sub) provides the following functions:

• Serial data communication

Interface type: RS-232C full duplex, DCE

Parameters: 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

Pin 1

Protective ground

Pin 2

Data in

Pin 3

Data out

Pin 7

Data ground

• Serial current loop

Interface type: 20mA current loop, full duplex

Parameters - 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

Pin 1

Protective ground

Pin 4

Data out (+)

Pin 5

Data in (+)

Pin 14 Data out (-)
Pin 16 Data in (-)