X troubleshooting – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual

Page 24

Advertising
background image

7. 631-X

TROUBLESHOOTING

Symptom

Possible Cause

Solution

Power Problems

Unit does not turn ON. Unit

turns on when power cord is
plugged in. LCD displays 000
when instrument is operating on
AC power.

Discharged battery or

Dead battery.

Recharge battery for a minimum of

14 hours. Refer to page 17.


Replace battery. Refer to page 21.

Unit does not turn on when

connected to AC power cord.

Open fuse.

Insufficient power.

Internal component failure.

Replace fuse. Refer to page 22.

Use a voltmeter to verify there is

power to the AC outlet.

Call AZI Customer Service for

information at 800-528-7411 or
602-470-1414.

Regeneration & Zero Problems

LCD displays .8.8.8.

Sensor saturated.

Do not attempt to adjust zero pot.

The sensor must be regenerated.
Refer to page 12 for information.

LCD displays .L.L.L when

taking first sample.

Changes in temperature.

Readjust zero pot. See page 13 for

information.

LCD displays H at finish of

sensor regeneration when zero is
pressed.

Internal contamination may

redeposit hydrogen sulfide from
flow system onto gold film
sensor.

Remove and replace fritware filter,

intake filter disk, scrubber filters
and Tygon tubing. Refer to “Flow
System” on Page 19.

Check tubing for kinks or crimps.
Repeat regeneration cycle. Refer to

page 12.

Zero adjust pot cannot be

adjusted to 0.

Pot not turned sufficiently.




Sensor may be ruptured or pot

may be broken.

1. Turn zero adjust up to 20 times

to reach the end. Pot will
“click” softly.

2. If no “0”, turn pot slowly in

opposite direction until display
reads “0”.

3. If still unchanged, call AZI

Customer Service at (800) 528-
7411 or 602-470-1414.

Page 24 of 50

AZI Customer Service 800-528-7411 or 602-470-1414