Fresh air solenoid – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual

Page 45

Advertising
background image

Fresh Air Solenoid


An external three-way solenoid can be used to provide fresh air or conditioned air during sensor
regeneration. This may be necessary if the sample stream lacks molecular oxygen. A low current
six volt DC solenoid, connected between pins 19 and 11 of the 25 pin rear panel connector, will be
energized during the regeneration cycle if the option board SW100 switch 6 is placed in the OFF
position.

If needed, the circuit may be built from the following components and configured as shown in the
following diagram. It will only function if the option board is installed in the 631-X instrument.

Required Parts:

Suggested Part

Similar AZI P/N

1 solenoid, 6volt 3way

Angar P/N 407569

1300 1004

1/8" to 1/16" tubing adaptor

Any

1300 0025

1/2" clamp, adhesive mount

Any

6000 0013

1/8" tube to instrument adaptor

Any

1400 3010

3" 1/8" tubing

Tygon Formula R3603

2500 3001

A/R 1/16" tubing

Tygon Formula R3603

2500 3002

1 25 pin male DB-25 connector
Solder-cup style

AMP 747912-2

None *

1 connector hood

AMP 749626-2

None *


* These are types not stocked by AZI, but should be available overnight from many AMP stocking
distributors such as Digi-Key Corporation. There are multiple suitable alternatives such as Radio
Shack’s 276-1547 and 276-1549.


TO SAMPLE
SOURCE

TO FRESH AIR

SOURCE

TO INSTRUMENT

1/8" TUBE

1/16" TUBE

1/16" TUBE

1/16" TUBE

PIN 1

PIN 1

REAR OF CONNECTOR,

CONNECT TO PINS 11 AND 19

MOST INSTRUMENTS

REAR OF CONNECTOR,

CONNECT TO PINS 19 AND 21

631-X SHIPPED IN 1994 ONLY

AZI Customer Service 800-528-7411 or 602-470-1414

Page 45 of 50