3.

Setup

 Place the pump on a stable surface.
 Plug the round connector end of the supplied power supply adapter into the power plug located on the

lower right of the pump's rear. See section 11, Logic Interface: TTL Input and Output, for a diagram
of the rear of the pump. Plug the other end of the power supply adapter into an appropriate electrical
outlet. The pump will be powered when the bottom of the power switch, located on the upper right of
the rear of the pump, labeled ‘1’, is pressed. The red indicator on the switch is visible when the power
switch is in the ‘on’ position. After power is applied to the pump, the pump’s display will flash.

 Next the Pumping Program can be entered. Before the Pumping Program can be operated, the pump

needs the measurement of the inside diameter, in millimeters, of the syringe that will be loaded. The
syringe diameter can be entered using the keypad on the front panel of the pump.

 Finally, the syringe can be loaded and the pump started.

4.

Loading Syringes

The syringe is loaded by securing the barrel and the pusher flange as follows:

1: Loosen the 2 thumbscrews on the syringe retainer bracket.

2: Press in fully the white drive-nut button on the pusher block, releasing the block. Taking care not to

drag the drive-nut on the drive-screw, slide the block away from the syringe holder, providing
sufficient space for the loaded syringe. Then release the white button.

3: Lift the syringe clamp above the syringe holder block. Turn it 1/4 turn and then lower it onto the

syringe holder block. The syringe clamp should be out of the ‘V’ slot.

4: Load the syringe with the barrel over the syringe holder and the syringe plunger towards the middle

of the pump. Place the barrel on the syringe holder block, in the ‘V’ slot, with the barrel flange
inserted between the syringe holder block and syringe retainer bracket.

5: On the pusher block, turn the thumbscrew to make the slot large enough for the plunger flange. Press

in fully the white drive-nut button on the pusher block, releasing the pusher block. Then slide the
block towards the syringe plunger. Place the syringe plunger flange into the slot and against the anti-
siphon plate
. When the flange is positioned in the slot, release the white drive-nut button.

6: Lift the syringe clamp to slightly above the height of the syringe barrel and turn the syringe clamp 1/4

turn back to its original position and then lower it onto the syringe barrel.

7: Firmly push in the syringe retainer bracket against the syringe barrel flange and tighten the 2

thumbscrews on the syringe retainer bracket. On the pusher block, turn the thumbscrew to move the
plate against the plunger flange.

 To unload the syringe, reverse the instructions for syringe loading.

5.

Guide Rod Collar Clamp

To protect a fragile syringe from damage caused by over infusion, use the collar clamp to limit the travel of
the pusher block. Using the hex wrench located in the tool holder on the rear of the syringe holder, loosen,
but do not remove, the hex screw on the guide rod collar clamp, enabling the collar clamp to slide on the
guide rod.

Position the collar clamp as required, and then tighten the hex screw on the collar clamp with the hex
wrench. Replace the hex wrench in the tool holder. When the pusher block comes in contact with the
collar clamp while infusing, a stall alarm will occur. The pump motor will be stopped and the Pumping
Program will be paused. If alarms are enabled, the buzzer will sound.