SP1000 series operator’s manual

Manual SP1000 series

39

10.4.1 Program Function Commands

The following commands are relevant to the currently select Program Phase. Note: During a Pumping
Program’s operation, the currently selected Phase can change automatically.
PHASE NUMBER

PHN [ <phase data> ]

Set/query currently selected Program Phase:

Set:
Currently selected Phase is set to <phase data>. Previous Phase is stored in non-volatile memory
and the requested Phase is recalled from the non-volatile memory. Set is only valid if the
Pumping Program is not operating.

Query response:

<phase data>

Currently selected Phase.

PUMPING PROGRAM FUNCTION

FUN [ <phase function> ]

Set/query the Pumping Program Phase’s function.

This command is relevant to the currently selected Phase. Set is only valid if the Pumping
Program is not operating

For a more detailed description of Program commands, see sec. 9.3, Program Function
Descriptions.

<phase function> =>

Rate Data Functions

When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.

RAT

Pumping rate. ‘RATE’

FIL

Fill syringe to dispensed volume. ‘FILL’

INC

Increment rate. ‘INCR’

DEC

Decrement rate. ‘DECR’

Non-Rate Control Functions

STP

Stop pump. ‘STOP’

PAS <number data>

Pauses pumping for ‘nn’ seconds. ‘PS:nn’

PAS <n.n>

Pauses pumping for 'n.n' seconds. 'PS:n.n'

PRI

Sub-Program Selection Input. 'Pr:In'

PRL <number data>

Sub-Program Start Label definition. 'Pr:nn'

LPS

Loop starting Phase. ‘LP:ST’

LPE

Loop end Phase. ‘LP:EN’

LOP <count data>

Loop to previous loop start ‘nn’ times. ‘LP:nn’

JMP <phase data>

Jump to Program Phase. ‘JP:nn’

IF

<phase data>

If Program input TTL pin low, jump to Phase. ‘IF:nn’

EVN <phase data>

Set event trigger trap. ‘EV:nn’

EVS <phase data>

Set event square wave trigger trap. 'ES:nn'

EVR

Event trigger reset. ‘EV:RS’

TRG <nn>

Override Operational Trigger Default Mode, set to <nn>,

‘tr:aa’
BEP

Sound short beep. ‘BEEP’

OUT <TTL level>

Set programmable output pin. ‘OUT.n’

PUMPING RATE

RAT [ C | I ] [ <float> [ <rate units> ] ]

Set/query pumping rate.