Troubleshooting and maintenance – Next Advance SP1070-M User Manual

Page 58

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1000 series operator’s manual

Manual SP1000 series

51

12.3.6 Foot Switch

Part #: ADPT-2

Allows the pump to be operated from a foot switch. Attaches to the TTL I/O connector.

12.3.7 Lockout Disable Key

Part#: ADPT-LOCKOUT-KEY

Enables setting Keypad Lockout mode and allows changing of settings while Keypad Lockout is set.

12.3.8 Firmware Upgrade

Part#: FW-1-NE1000

Upgrades to the newest version of the firmware.

Part#: FW-1X-NE1000

Gradient/Linear ramping for smooth gradients, plus other new program

functions.
Part#: FW-1X2-NE1000 Also expands the Pumping Program memory to 340 Phases.

Contact your dealer for these upgrades and to determine the current available version of the pumps internal
firmware.

12.4

Troubleshooting and Maintenance

Maintenance: Periodically, a small amount of all-purpose oil should be applied to the guide rods and
drive screw.

The mechanism should be kept clean to prevent impeded operation.

No other special maintenance or calibrations are needed

RS-232 Communications: If no RS-232 communications is possible or garbled responses are received
from the pump, check the following:

If the triangle appears in the upper left of the LCD display, then the pump is receiving valid
communications. The communications problem is probably with the receiving communication application
or with the receive line on the cable.

If the Basic communications mode is used, check if the pump is in Safe communications mode. See
section 10.2, RS-232 Protocol:, for instructions on how to change the communications mode.

Verify the pump’s baud rate and network address. To set the RS-232 communications parameters, see
section 6.12, ‘Setup’ .

Using a lower baud rate may also improve the reliability of the RS-232 communications.

Pusher block makes a snap or click sound when the pump is started: This is a normal condition.
When the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive
screw. The sound heard is the drive-nut engaging on the drive screw.

Pump doesn’t stop after dispensing a set volume: The pump was previously setup with a multiple Phase
Pumping Program. To simply dispense a fixed volume at a fixed pumping rate, the second Program Phase
must be the ‘Stop’ function. See section 9.1, How to Enter Pumping Programs, for instructions on
changing the Pumping Program.

Pump stops pumping after a period of time: A dispense volume has been set. Verify that the dispense
volume is set to 0.

Advertising
This manual is related to the following products: