SP1000 series operator’s manual

Manual SP1000 series

i

Quick Start Instructions

Syringe Pump General Operating Instructions

This document is meant to be a quick guide to use for operating Next Advance Syringe Pumps. For more
detailed operating instructions, please see the full user manual. If you have any additional questions, please
contact us at [email protected]

Note- To reset the pump, press and hold the right-most arrow key while turning the power on.

Loading the Syringe
-Press in the white drive-nut button to move the pusher bar
-Insert the syringe plungers in the pusher bar bracket slots.
-Insert the syringe barrel flange in the flange in the flange bracket slots with the syringe barrel holder
lowered onto the syringes. Tighten the flange bracket onto the syringe flanges.

Changing Numerical Values on the Display
-Each of the four arrows corresponds to a digit on the display. To change the decimal place, hold down the
left most key until it is located where you need it to be.
-Whenever you enter a new value, wait 2 seconds for the display to blink – this means that the new value is
stored in memory.

Basic Programming
Rate- Press the “Rate” button to enter the rate at which you would like the syringe to infuse or withdraw.
An or:01 error means that the pumping rate is out of range. The current units are indicated by the red lights.
To change units, press the rate key twice. UNtS will appear in the display. Press any arrow key to change
units. Rate can be changed while the unit is pumping.

Volume- Press the “Volume” button to enter the amount of liquid you would like for the syringe to
withdraw or infuse. Pressing the “Volume” button twice will display the volume that has been dispensed
so far (the “Dispensed” light will come on). To reset “volume dispensed” value, press and hold any arrow
key while it is displayed, change the syringe diameter, or turn off the pump.
If the volume is set to 0, the display will read “OFF”- this indicates that the instrument is set for continuous
pumping.
Default units are set according to the inside syringe diameter (diameters set from 0.1-14.0 mm will allow
you to set volume units in µl, while diameters set from 14.01-50.00 mm will allow you to set volume units
in ml). TO OVERRIDE DAFAULT VOLUME UNIT SETTINGS- Change the syringe diameter (this is
the required first step). Press the “Volume” button twice to view volume dispensed. Press any arrow key
and “UntS” will be displayed. Press arrow keys to toggle between ml and µl. Changing the diameter will
not affect volume units until again until the system is reset.

Diameter- The inside diameter range of syringes that can be used in the pump is 0.1mm to 50.0mm. Press
the diameter key once to set the value.

Direction- Syringe will be in infuse mode, until “→→←←“ button is pressed and the “withdraw” light is
lit. Direction can be changed while the unit is pumping if the “Volume to be Dispensed” is 0.0 (off).
NOTE: This parameter is not available in the SP300 “Just infusion” syringe pump.

Start/Stop- Press “Start/Stop” button to start pump. While pump is running, pressing the “Start/Stop”
button will put the pump in pause mode (the “pumping” light will blink). Pressing the “Start/Stop” button
again will cause the unit to resume the following the program. To start program over, shut down the power
(the program will not be lost until the pump is reset). Holding down the “Start/Stop” button will cause the
unit to purge the liquid in the syringe.

Parameter Set up
To change various parameters related to the general operation of the syringe pump, it is necessary to enter
into “Setup” mode in order to view all of the options.
To enter “Setup” mode, press and hold down the “Diameter” button. Press any non-arrow key to toggle
through selections. For parameters followed by a number, 0=off and 1=on.