Next Advance SP1070-M User Manual

Page 35

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1000 series operator’s manual

Manual SP1000 series

28

Phase

Function

Rate

Volume

Direction

7

RATE

300.0 ml/hr

1.5 ml

Infuse

Phase

Function

8

BEEP

Phase

Function

9

LP:ST

Phase

Function

10

PS:60

Phase

Function

Rate

Volume

Direction

11

RATE

500.0 ml/hr

3.75 ml

Infuse

Phase

Function

12

LP:03

Phase

Function

Rate

Volume

Direction

13

RATE

900.0 ml/hr

17.25 ml

Withdraw

Phase

Function

14

BEEP

Phase

Function

15

PS:00

Phase

Function

16

LP:EN

9.4.5 Example 5: Control from a High-Low Pressure Sensor

10

25

m

l/h

r

Pumping
Rate

Program
Output Pin

Event Input
Pin

Pressure high sensor

Pressure low sensor

Pressure trigger points

This example demonstrates a Pumping Program whose control depends on an external sensor. Assuming a
pressure sensor that is configured to detect a high pressure point and a low pressure point, the Pumping
Program individually selects whether it will react to the high or low pressure point.

The “Program Output” pin on the TTL I/O connector (pin 5) is used to select the high or low pressure
point. When low, the low pressure point is selected (PH:01), and when high, the high pressure point is
selected (PH:05). The Program begins by infusing continuously at 10.0 ml/hr (PH:02), while a background
trap is set for the low pressure point (PH:03). To create a delay when the pressure sensor is switched from
high pressure to low pressure when the “Program Output” pin is set, a small volume is pumped (PH:02, 06)
before the background traps are set.

When the low pressure trap is triggered, the pump sets the high pressure trap (PH:07) and begins to
increment the flow rate. The flow rate is incremented in 1.0 ml/hr steps with every 0.25 ml dispensed
(PH:08-10). If the high pressure trap hasn’t as yet been triggered, the flow rate will max out at 25.0 ml/hr
while waiting for the high pressure trap (PH:11). When the high pressure point is reached, the pump
immediately will drop down to 10.0 ml/hr (PH:02), and once again wait for the low pressure point.

Advertising
This manual is related to the following products: