Next Advance SP1070-M User Manual

Page 39

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1000 series operator’s manual

Manual SP1000 series

32

The pump then withdraws 0.25 ml (PH:07), pauses for 1 second (PH:08), then repeats this process if the
Program Input TTL line is low (PH:09), otherwise it continues with the next Phase.

Next, the pump pauses for 10 seconds (PH:10). Then it pauses again for the lesser of another 10 seconds
(PH:12) or until an Event Trigger occurs (PH:11). The Program then restarts (PH:13).

Phase

Function

1

ET:RS

Phase

Function

2

OUT.1

Phase

Function

Rate

Volume

Direction

3

RATE

800.0 ml/hr

5.0 ml

Infuse

Phase

Function

4

OUT.0

Phase

Function

5

ET:07

Phase

Function

Rate

Volume

Direction

6

RATE

800.0 ml/hr

0.0 ml (OFF)

Infuse

Phase

Function

Rate

Volume

Direction

7

RATE

1000.0 ml/hr

0.25 ml

Withdraw

Phase

Function

8

PS:01

Phase

Function

9

IF:07

Phase

Function

10

PS:10

Phase

Function

11

ET:01

Phase

Function

12

PS:10

Phase

Function

13

JP:01

9.4.9 Example 9: Automatic Refill

While a foot switch is held, dispense continuously, starting with a 1.5 mL bolus. When the foot switch is
released, the syringe will refill to the volume that was dispensed.

In the pump’s general setup, set the TTL Trigger to “Foot Switch Hold” mode (TR:FH).

Phase

Function

1

TR:ET

Redirect trigger input to the Event Trap

Phase

Function

2

ET:05

Set Trigger Event Trap to start syringe refill

Phase

Function

Rate

Volume

Direction

3

RATE

1000.0 ml/hr

1.5 mL

Infuse

Begin bolus dispense

Phase

Function

Rate

Volume

Direction

4

RATE

500.0 ml/hr

0.0 ml (OFF)

Infuse

Then infuse continuously

Phase

Function

Rate

5

FILL

1000.0 ml/hr

Refill syringe when foot switch is released

Advertising
This manual is related to the following products: