AGM300 - Connecting External Alarms

Overview

The AGM300 includes 4 Form C SPDT alarm contacts for the connection of external alarm devices. The relays
are factory assigned to the following conditions:

Relay

#1 Leak

Relay

#2 Spill

Relay

#3 Evacuate

Relay

#4 Fault

Connection

Use the AC conduit or any of the remaining service knockouts to gain cable access to the interior of the
monitor. Locate the relay connector (Page 7) and remove it from the circuit board. Secure the leads to the
connector orienting them as shown in the diagram below. When you are through securing the connections,
carefully plug the connector back onto the circuit board.

AGM300 Relay Connector

NOTE: Each relay may be connected as normally open (NO), or normally closed (NC).

Latching / Non-Latching – refer to Page 38 to select Alarm Ack Mode.

Alarm power source – power for the alarm devices connected to the relay contacts may be supplied from the
Auxiliary AC Power Out Connector, 5 amp. max. (Page 7).