Bacharach AGM300 User Manual

Page 83

Advertising
background image

Instruction 3015-4275

75

Cal Data Register

Register 0x0014h

R/W

174 bytes

Variable Type

Length

Description

Factor

Float

132 bytes

33 element array containing cal factors for each of the 33
gases. Order of Cal Factor is defined in Note 1 under Zone
Data

Curve

UC

3 bytes

Existing Master Curve to use for user defined “NEW GAS”

New Gas Name

C

39 bytes

3 x 13 array that contains the names for each of 3 new
gases

NOTE: “Future” Gases are gases that will be added to the library at a later date. They will require generation
of new cal curves that will be added to the code. “New” Gases can be added in the field by selecting a similar
existing cal curve and then defining a cal factor that will bring the accuracy within the desired range. It is
strongly recommended that any New Gas curves that are added be done in conjunction with help from
Bacharach, Inc.

Date Time Register

Register 0x0015h

R/W

14 bytes

Variable Type

Length

Description

Date_Time

Time

14 bytes

Contains current time and date. Structure is defined as in
note 2 of zone data

Time Structure Format consists of 14 unsigned character types. They are 1 second digit, 10 second digit, 1
minute digit, 10 minute digit, 1 hour digit, 10 hour digit, a day digit, 10 day digit, 1 month digit, 10 month digit, 1