Bacharach AGM300 User Manual

Page 79

Advertising
background image

Instruction 3015-4275

71

System Data Register

Register 0x0010h

R/W

54 bytes

Variable Type

Length

Description

Type

UI

2 bytes

Indicates EEPPROM has been initialized if value = 300
DO NOT MODIFY

REV

Float

4 bytes

Firmware Rev Level DO NOT MODIFY

SN

UI

2 bytes

Firmware Serial Number DO NOT MODIFY

Node

UC

1 byte

Network Slave Node # (valid values are 1-15). The default
is that indicated by the Node DIP Switch on main board.

Location

C

13 bytes

Array defining text name of unit

Stop_Bits

C

1 byte

Number of stop bits used in the AGM300 data stream.
Default = 1. Other available value is 2

Aud_Alarm

UC

1 byte

Sounds internal board buzzer on condition; 0 = no buzzer;
1= Alarm; 2=Fault; 3=Leak; 4=Spill; 5=Evacuate; 6=Monitor
Off line (DEFAULT = 0)

Alarm_Ack_
Mode

UC

1 byte

Defines Alarm Operation. Manual Acknowledge = 0; Auto
Acknowledge = 1 (DEFAULT = 0)

Num_Zones

UC

1 byte

Number of install zones (Value initialized during auto
detect during Power On Self Test).

UNUSED TIM

13

bytes

UNUSED

Rezero_Mode

UC

1 byte

Defines rezero mode. Auto Rezero = 0; Rezero every
zone = 1 (DEFAULT = 0)

Hold_Time

UI

2 byte

Length of zone hold interval in minutes (DEFAULT = 15
minutes)

Det_Limit

UC

1 byte

Minimum detection limit (in PPM). Concentrations less
than or equal this value will read as 0 PPM (DEFUALT = 0
PPM)

Avg_Size

UC

1 byte

Size of running average used in computing PPM value. DO
NOT MODIFY
.

Loop2_factor

Float

4 bytes

Defines PPM current loop output. (DEFAULT =
0.16mA/PPM)

Serv_Mode_TO

UI

2 bytes

Service Mode Timeout value (in minutes). (DEFAULT = 60
MINUTES)

RS-485_BAUD

UI

2 bytes

BAUD RATE for RS-485 connection (between ADM800
and AGM300 or MODBUS master and AGM300 depending
on the Network topology). Default=9 (19.2K); other values
are 8=9600, 7=4800

UNUSED UI

2

bytes

UNUSED