Instruction 3015-4275

31

Display
showing
“WARM UP”

ADM800 - Screen Displays

When first powered up a splash screen appears indicating the serial number and firmware revision number.
After a brief moment the System Screen is displayed. Please refer to the Appendix for a logic diagram of all
the screen displays.

There is a 15-minute warm-up period after power is first applied. Under “AGMS”, "WARM UP" is displayed.
The green indicator on the AGM will blink during the warm-up period.

IMPORTANT: DO NOT plug the gas-sample hose or inlet port of any zone. The monitor will perform a vacuum
test during start up and a plugged zone will give a false indication.