Zone Data

Register 0x12xxh R/W 78 bytes

Each zone for an AGM300 has a separate Zone data structure that is 78 bytes long. The zone number is the
low order byte in the register address (i.e., Zone 1 data register = 0x1201h)

Variable Type

Length

Description

Location

C

13 bytes

13 byte array, Alpha Numeric Description or Name of Zone

Flow OK

UC

1 byte

Status of Flow check. Value of 1 indicates flow check is
good. DO NOT MODIFY

Refrig. Type

UC

1 byte

See note 1 Below (DEFAULT = R134a)

Distance

UI

2 bytes

Zone Tubing Length (in feet) (DEFAULT = 100 feet)

Zone Temp

I

2 bytes

Avg temp at zone (degrees C) (DEFAULT = 25

°C)

Concentration

Float

4 bytes

Last Measured concentrations (uM/L) DO NOT MODIFY

Concentration2

Float

4 bytes

Last Measured concentration (PPM) DO NOT MODIFY

Alarm Ack

UC

1 bytes

Set value to 1 to acknowledge Alarm. NOTE: AGM300 will
reset this byte to 0 when the Alarm byte (below) is = 0 and
zone in alarm is sampled. If the alarm condition/byte
increases (leak>>spill or spill>>evac) the AGM300 will also
reset this byte to 0

Alarm

UC

1 bytes

Alarm Status; 0 = no alarm, 1 = leak; 2 = spill; 3 =
evac.

Leak Level

UI

2 bytes

Level to trigger a leak alarm (in PPM) (DEFAULT = 100)

Spill Level

UI

2 bytes

Level to trigger a spill alarm (in PPM) (DEFAULT = 300)

Evac Level

UI

2 bytes

Level to trigger a evacuate alarm (in PPM) (DEFAULT =
500)

Peak PPM

UI

2 bytes

Highest Recorded PPM in zone

Peak Time

TIM

13 bytes

Date an time of highest peak (see note 2 for format)

Alarm Time

TIM

13 bytes

Date and time of last alarm (see note 2 for format)

Not Used

13 bytes

Log Interval

UI

2 bytes

Number of minutes between Log entries (DEFAULT =
1440)

NOTE 1: Refrigerant Types Defined from 00h to 20h in the following order : 00h=R11, 01h=R12, 02h=R22,
03h=R23, 04h=R113, 05h=R114, 06h=R123, 07h=R124, 08h=R134a, 09h=R401a, 0Ah=R402a, 0Bh=R402b,
0Ch=R404a, 0Dh=R407a, 0Eh=R407c, 0Fh=R409a, 10h=R410a, 11h=R500, 12h=R502, 13h=R503,
14h=R507, 15h=R508b, 16h=H1301, 17h=R408a, 18h=Future2, 19h=Future3, 1Ah=Future4, 1Bh=Future5,
1Ch=Future6, 1Dh=Future7, 1Eh=New1, 1Fh=New2, 20h=New3

NOTE 2: Time Structure Format consists of 13 unsigned character types. They are 1 second digit, 10 second
digit, 1 minute digit, 10 minute digit, 1 hour digit, 10 hour digit, 1 day digit, 10 day digit, 1 month digit, 10 month
digit, 1 year digit, 10 year digit, last byte is unused.