In the window’s left part there are three columns: in the first one it is possible to

insert the instrument’s output voltage (i.e. the abscissa values of the calibration

curve), while in the second one it is possible to insert the temperature/pressure

value corresponding to the previous voltage (i.e. the ordinate values of the

calibration curve). These values will be interpolated using a polynomial. In the

third column it is reported the “Curve Error”, i.e. the difference between the

computed curve and the experimental values.

To set how many experimental values you wish to use to create the calibration

curve, please place a check in the column besides the abscissa values.

In the window’s middle part it is reported both the sensor’s calibration curve and

the 5 coefficients (from a

0

to a

4

) of the interpolation polynomial. Please, use

button “Compute Curve” to refresh the calibration curve to apply any changes.

On the right of the calibration curve there are some boxes, where you can insert

the sensor name, the desired measure units and the sensor type (Temperature,

Pressure or Other).

Below these boxes there are some pushbuttons which allow the user to save or

delete a customized sensor.

In the window’s upper right corner you can see some default “custom” sensors. If

you wish to load one, please click twice on the sensor’s name.

Once correctly defined the new sensor, the user may exit Customization sensor

mode, by clicking on the OK pushbutton, and has to set the new sensor on the

desired channel (press button Channels and follow the instructions reported in the

channels paragraph).