Control Company 4048 TRACEABLE MEMORY MONITORING THERMOMETER WITH ALARM User Manual

Control Company Thermometers

Advertising
background image

TRACEABLE MEMORY MONITORING

THERMOMETER WITH ALARM INSTRUCTIONS

CONTROL COMPANY

4455 Rex Road • Friendswood, Texas 77546 USA

Ph. 281-482-1714 Fax 281-482-9448

E-mail [email protected] • www.control3.com

Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025 accredited as a Calibration Laboratory by A2LA.

Cat. No. 4048/4127

Traceable® Memory Thermometer with Alarm Instructions 92-4048-00

Rev. 2 121008

Traceable® is a registered trademark of Control Company

© 2008

Control Company

SPECIFICATIONS

Range:

–58 to 158.0°F / –50 to 70.0°C

Resolution: 1°

Accuracy: ±1°C

Sampling

Fast 10 seconds or Normal

Rate:

60 seconds (user selectable)

Features:

LCD digital thermometer, internal or external display with maximum/minimum

memory display and high/low alarm

Battery:

AA Alkaline

PROBES

4048 Model Probe:

A detachable probe sensor and ten feet of cable are supplied with

the unit. A wire probe holder is provided for permanently mounting

the probe. To mount the probe, place a nail or screw through the

hole of the wireholder and secure the holder to a surface.

4127 Model Probe:

Probe is sealed in a miniature bottle (1 x 2½ inches) filled with a

patented nontoxic glycol. Solution is GRAS (generally recognized

as safe) by the FDA (Food and Drug Administration). It eliminates

concerns about incidental contact with food or drinking water. The

solution filled bottle simulates the temperature of other stored

liquids. Velcro

®

and magnetic strip are supplied to mount the bottle

to the inside of a refrigerator/ freezer and to mount the display unit

to the outside. Micro-cable permits refrigerator doors to close on it.

NOTE: Plug probe sensor into the jack on the side of the unit. Press RESET button once

whenever probe sensor is removed or inserted. Updated ambient or probe temperatures will

be displayed.
With probe sensor attached the unit simultaneously displays current probe temperature and

probe minimum/maximum temperatures.
With the probe sensor removed the unit simultaneously displays current ambient tempera-

ture and ambient minimum/maximum temperatures.
RESETTING THE UNIT

Use a pointed object to push the RESET button on the back of the unit. It is necessary to

reset the unit whenever a change is made to

o

C/

o

F temperature , FAST/NORMAL sampling

rate, or the probe sensor is plugged in or removed.
NOTE: HI/LO temperature alarm settings and MIN/MAX memories will be cleared after

resetting the unit.
OPERATION MODES

The unit has two operating modes: NORMAL DISPLAY MODE and ALARM DISPLAY

MODE. To change from one mode to another press the MODE button once.
NORMAL DISPLAY MODE Indicated by the display of current temperature and the

MIN (minimum) and MAX (maximum) temperatures. These

temperatures relate to external (probe) temperature if the

probe is plugged in and to internal (ambient) temperature if

the sensor probe is removed.

ALARM DISPLAY MODE

Indicated by the display of current temperature and the HI

and LO alarm settings. These temperatures relate to external

(probe) temperature if the probe is plugged in and to internal

(ambient) temperature if the sensor probe is removed.

RESETTING MIN/MAX MEMORIES

There are two ways to reset the memories:

1. To reset both minimum and maximum memories - press MEMORY CLEAR button.

2. To reset individual minimum and maximum memories - press either the MIN or MAX

button (on back of unit) once to reset the respective memory. The other memory will

remain unaffected.

NOTE: You must be in Normal Display Mode to reset the memories.
DISPLAYING °F OR °C

To display the temperature readings in Fahrenheit (°F) or Celsius (°C) slide the switch on

the back of the unit to °F or °C.

NOTE: The unit must be reset after changing the temperature unit.
SELECTING THE SAMPLING RATE

To select normal sampling rate (60 seconds) or fast sampling rate (10 seconds) slide the

switch on back of unit to NORMAL or FAST.

NOTE: The unit must be reset after changing the sampling rate. Using the NORMAL

sampling rate will extend battery life.
SETTING THE HIGH/LOW ALARM

1. Press the MODE button until unit is in the ALARM DISPLAY MODE indicated by the

display of current temperature and HI and LO alarm settings.

2. To set the HIGH temperature alarm limit press the HI button, to advance the display in

1° increments, until the desired high temperature is displayed. Press and hold the HI

button to rapidly advance the display in 5° increments.

3. To set the LOW temperature alarm limit press the LO button, to advance the display in

1° increments, until the desired low temperature is displayed. Press and hold the LO

button to rapidly advance the display in 5° increments.

ALARM

The unit will sound the alarm in both NORMAL DISPLAY MODE and ALARM DISPLAY

MODE.

Slide the ALARM ON/OFF switch to ON. When the current temperature display rises

above the HI set point, or falls below the LO set point, the alarm will sound for 1 minute

and issue a 3 second repeater beep every minute thereafter, for up to 12 hours. The

alarm will continue to sound even if the current temperature display returns to an “in

range” condition.

To temporarily silence the alarm press either the HI or LO button the back of the unit.

The unit is still active and the alarm will sound again if the temperature goes out of range.

To permanently silence the alarm slide the ALARM ON/OFF switch to OFF position.
BENCH STAND

The unit is supplied with a bench stand that is a part of the battery cover. To use the

bench stand, locate the small rectangular opening at the bottom of the unit. Place your

fingernail into the opening and flip the stand up. To close the stand, simply snap it shut.
WALL MOUNTING THE UNIT

Set a screw into the wall at the desired location. Do not set the screw flush to the wall,

the head of the screw will need to slip into the receptacle on the back of the unit. Once

the screw has been properly set, hang the unit in place by sliding the receptacle on the

back of the unit over the head of the screw.
ALL OPERATIONAL DIFFICULTIES

If this thermometer does not function properly for any reason, please replace the battery

with a new high-quality battery (see “Battery Replacement” section). Low battery power

can occasionally cause any number of “apparent” operational difficulties. Replacing the

battery with a new fresh battery will solve most difficulties. If the voltage of the battery

becomes low °C and °F symbols will flash.
BATTERY REPLACEMENT

Erratic readings, a faint display, or no display are all indications that the battery must be

replaced. Slide the battery cover toward the end of the unit. Remove the exhausted bat-

tery and replace with an AA alkaline battery (Cat. No. 1111). Replace the battery cover.

WARRANTY, SERVICE, RECALIBRATION

For warranty, service or recalibration contact:

Advertising