The sampling rate which can be set is also
dependent on the active measurement
channels:

• 0.1 ms

max. 2 analog input

channels

• 0.2 ms

max. 4 analog input

channels

• 0.5 ms

all 10 input channels

• 1.0 ms

for SMART sensors

Recording time

A drop-down list appears when this item is
selected. Using the arrow keys (

↑) and (↓)

you can select a recording time of between 1
second and 24 hours, followed by pressing
OK

.


When changing the setting, information is
shown at the bottom of the display pertaining
to the recorded data showing you the result
of your setting.

Recording option

The options available here are End when
time has elapsed

and Record

continuously until terminated

.


End when time has elapsed

means that

recording is discontinued when the time
entered in Recording time has elapsed.

Record continuously until terminated

causes the oldest measured values to be
deleted when new measured values are
added upon the time entered in Recording
time

elapsing. The time entered in

Recording time

remains as the "past" in

relation to the current time in the measured
value memory until the Stop command is
given (Stop + OK or actuate the Esc key).