6.5 Technical Specifications


Sensor Inputs


The HMG features:

• 4 sensor input jacks (channel A – H) for up to 8 analog sensors,

or up to 4 SMART sensors

• 1 input jack with 2 digital inputs (channel I - J) and

a voltage input of -10 V to + 10 V (shown at channel H)

Connecting the sensors is done using an standard M12x1 plug connector (5 pin).

Channel A to H:

• Automatic detection for HSI sensors (pressure, temperature, volumetric flow

rate transmitters and SMART sensors)

• Connection of standard sensors with current or voltage signals
• Differential channels for channel A - B; channel C - D; Differential channel for

flow rate measurement orifice (shown at channel B)


Channel I and J:

• Frequency channels (speed (rpm) measurement, counting function)

Frequency range: 1 – 30,000 Hz

(

≤ ± 0.1% FS max.)

Switching threshold / switch-back threshold: 2V/1V
Max. input voltage: 50 V; display delay max 2 sec.

Channel A and B (accuracy)

Channel C and D (accuracy)

4 – 20 mA

(

≤ ± 0.1% FS max.)

4 – 20 mA

(

≤ ± 0.1% FS max.)

0 – 20 mA

(

≤ ± 0.1% FS max.)

0 – 20 mA

(

≤ ± 0.1% FS max.)

0 – 10 V

(

≤ ± 0.1% FS max.)

0 – 10 V

(

≤ ± 0.5% FS max.)

0 – 5 V

(

≤ ± 0.2% FS max.)

0 – 50 V

(

≤ ± 0.1% FS max.)

1 – 5 V

(

≤ ± 0.2% FS max.)

0 – 5 V

(

≤ ± 1% FS max.)

1 – 6 V

(

≤ ± 0.2% FS max.)

1 – 5 V

(

≤ ± 1% FS max.)

0.5 – 4.5 V

(

≤ ± 0.1% FS max.)

1 – 6 V

(

≤ ± 1% FS max.)

0.5 – 5.5 V

(

≤ ± 0.2% FS max.)

0.5 – 4.5 V

(

≤ ± 0.1% FS max.)

0.5

5.5

V (

≤ ± 1% FS max.)

Channel E to G (accuracy)

Channel H (accuracy)

4 – 20 mA

(

≤ ± 0.1% FS max.)

4 – 20 mA

(

≤ ± 0.1% FS max.)

0 – 20 mA

(

≤ ± 0.1% FS max.)

0 – 20 mA

(

≤ ± 0.1% FS max.)

0.5 – 4.5 V

(

≤ ± 0.1% FS max.)

-10 – +10 V

(

≤ ± 0.5% FS max.)

0.5

4.5

V (

≤ ± 0.1% FS max.)

Battery service times

(battery is fully charged)

• HMG 3000 without sensors

ca. 11 hours

• HMG 3000 with 2 sensors

ca. 9 hours

• HMG 3000 with 4 sensors

ca. 7 hours

• HMG 3000 with 8 sensors

ca. 4 hours

Advertising