HL, SL

17


Caution:

The temperature controlling i.e. of fluids in a reactor constitutes normal circulator
practice.

We do not know which substances are contained within these vessels.
Many substances are:
• inflammable, easily ignited or explosive
• hazardous to health
• environmentally unsafe
i.e.: dangerous

The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce the
risks to a minimum.
• Are all tubes and electrical cables connected and installed?

Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.

• Do dangerous steams or gases arise when heating?

Is an exhaust needed when working?

• What to do when a dangerous substance was spilled on or in the unit?

Before starting to work, obtain information concerning the substance and
determine the method of decontamination.


Notice:

Check the safety installations at least twice a year!

• Excess temperature protection according to IEC 61010-2-010.

With a screwdriver turn back the adjustable excess temperature protection until the
shut-down point (actual temperature).

• Low level protection according to IEC 61010-2-010.

To check the function of the float, it can be manually lowered with a screwdriver for
example.